H10: manufacturing review: a review of manufacturing records was not performed, as additional complaints have not been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation and image was not provided for review.The proximal end of the stent was placed in a straight section prior to deployment.Based on the investigation of the provided information, the investigation is closed as inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding the preparation of the device the instructions for use states 'flush the stent graft lumen with sterile saline by using a small volume syringe.A) attach the syringe to the luer port at the back of the endovascular system (.).B) attach the syringe to the luer port on the y-adapter, (.) close the stopcock when flushing is complete and remove the syringe from the luer port'.The instructions for use states regarding the accessories 'the use of an appropriately sized introducer sheath is recommended; prepare a stiff 0.035" guidewire per its instructions for use and advance the guidewire under fluoroscopy to the target location' regarding the precautions prior and during deployment the instructions for use states 'prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure' and 'if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.' the packaging pictograms indicate an introducer size of 10f and a 0.035" guidewire.H10: d4 (expiry date: 03/2024), g3.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|