MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Medical Device Problem Codes Entrapment of Device (1212); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Eye Injury (1845)
Date of Event 08/04/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that the intraocular lens got trapped in the incision during implantation of the intraocular lens (iol).The surgeon advanced the lens and noticed a fracture in the optics.The incision was widened.The iol was removed and discarded.A back up lens was implanted without harm to the patient.Additional information has been requested; however, further information has not been received.
 
Additional Manufacturer Narrative
The product was returned for analysis and damage was observed to the iol (intraocular lens).No cartridge was returned.Iol returned pressed down into well area of iol case base.Solution is dried on both surfaces of the optic and haptics.One haptic is nicked/chipped-rejectable near the gusset area, the other is bent/twisted.The optic is cracked/fractured almost dividing the iol in two and scratched/marked-rejectable.Unable to conduct dimensional testing due to condition of returned sample.The root cause for the reported complaint "trapped in the incision, fracture of the optics, incision widened, in and out" could not be determined.Only the iol was returned for evaluation.No cartridge was returned.Due to this, we are unable to complete a thorough investigation to determine a root cause.The reported complaint is lacking in relevant information such as associated products used to complete a thorough investigation.Due to the lack of information, we are unable to verify if the iol contributed to the event.Unable to conduct dimensional testing due to condition of returned sample.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key12364715
Report Number9612169-2021-00172
Device Sequence Number3974190
Product Code HQL
UDI-Device Identifier00380655093290
UDI-Public00380655093290
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/10/2026
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.250
Device Lot Number25130407
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/09/2021
Supplement Date Received by Manufacturer07/18/2022
Initial Report FDA Received Date08/25/2021
Supplement Report FDA Received Date08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-