Model Number 1094-04
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via medwatch (b)(4) that the pigtail catheter tip broke and was snared to remove.A percuflex guide catheter was selected for use in ureteral stent placement (stent from kidney to bladder).During the procedure, upon placing a boston scientific right ureteral stent system, it was noted that the pigtail tip located in the kidney broke.The physician was able to snare the broken loop and the procedure was completed after snaring the broken piece of the pigtail catheter.Imaging was reviewed indicating that the ureteral stent was functioning appropriately at the end of the case.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUFLEX
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12366128
Report Number2134265-2021-10740
Device Sequence Number4633229
Product Code FAD
UDI-Device Identifier08714729170556
UDI-Public08714729170556
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K924608
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/03/2024
Device Model Number1094-04
Device Catalogue Number1094-04
Device Lot Number0026913303
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2021
Initial Report FDA Received Date08/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-