Model Number FDS50015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Stroke/CVA (1770)
Date of Event 07/29/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device not returned for evaluation.
 
Event or Problem Description
It was reported during the procedure the patient suffered a stroke with unknown relation to the stent (subject device).The patient was placed on a ventilator and had a nih score of 18 after the event.No further information is available.
 
Additional Manufacturer Narrative
The subject device is not available; therefore, visual and functional testing as well as physical analysis cannot be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Due to the automated mes system there are controls in the manufacturing process to ensure the product met specifications upon release.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.It was reported that after deploying the subject stent the patient had a stroke.The patient is currently on a ventilator, nih 18 last follow up.Based upon medical review, the harm observed in the complaint is anticipated in nature as per the device risk assessment.As a product related root cause does not apply and the issue is due to a known physiological effect of the procedure and/or patient condition noted with the directions for use, product labeling and/or risk documentation files, an assignable cause of anticipated procedural complication will be assigned to this complaint.H3 other text : device not returned for evaluation.
 
Event or Problem Description
It was reported during the procedure the patient suffered a stroke with unknown relation to the stent (subject device).The patient was placed on a ventilator and had a nih score of 18 after the event.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURPASS EVOLVE 5.0MM X 15MM - PMAS
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key12366355
Report Number3008881809-2021-00346
Device Sequence Number14408428
Product Code OUT
UDI-Device Identifier07613327386691
UDI-Public07613327386691
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P170024/S003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberFDS50015
Device Catalogue NumberFDS50015
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/29/2021
Supplement Date Received by Manufacturer11/04/2021
Initial Report FDA Received Date08/25/2021
Supplement Report FDA Received Date12/02/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AXS CAT 5 CATHETER (STRYKER).; AXS CAT 5 CATHETER (STRYKER)
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
Patient SexFemale
-
-