• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-55
Device Problem Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/06/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: a getinge field service engineer (fse) was dispatched to investigation.The fse evaluated the iabp unit and observed that the reported issue involved the power supply board.At this time, repairs are ongoing and a supplemental report will be submitted when additional information is provided.The full event site name in is (b)(6).
 
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) was not working on mains power.It is unknown under which circumstances the event occurred or if a patient was involved; however there was no adverse event reported.
 
Manufacturer Narrative
Updated data: b4,e1(email),e2,e3,g2,g3,g6,g7,h2,h10,h11.Corrected data: b5,d1,d5,d10,h6(impact).
 
Event Description
It was reported that during routine check, the cs300 intra-aortic balloon pump (iabp) was not working on mains power.There was no patient involvement.
 
Manufacturer Narrative
Updated fields: e1(site country), h6 (,type of investigation, investigation findings, component codes, investigation conclusions).E3 is unknown (initially it is submitted as "other healthcare professional").A getinge field service engineer (fse) evaluated the iabp unit and was able to reproduce the reported issue.To fix the issue, the fse replaced the power supply, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released for cleared for clinical service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12366644
MDR Text Key268106358
Report Number2249723-2021-01899
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567112541
UDI-Public10607567112541
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-55
Device Catalogue Number0998-00-3023-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-