| Model Number |
1012454-20 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/10/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a lesion in the heavily calcified right coronary artery.A 4.5x20mm nc trek balloon dilatation catheter (bdc) was advanced and used without issue.However, the bdc felt resistance during removal with an unspecified guiding catheter but was removed successfully to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection was performed on the returned device.The reported difficulty removing the device with the guiding catheter could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|