• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ULTRAFLEX SELF-ADHERING MALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ULTRAFLEX SELF-ADHERING MALE EXTERNAL CATHETER Back to Search Results
Model Number 33301
Medical Device Problem Code Inadequate or Insufficient Training (1643)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.A potential root cause could be "user not familiar with product".It was unknown whether the device had met specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: intended use: the self-adhering male external catheter is used for the drainage of urine.The catheter is applied by the patient or caregiver.Description / indication: the self-adhering male external catheter is designed for the management of adult male urinary incontinence.Contraindication: do not use on irritated or compromised skin.Warning: reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury or illness of the patient.Precaution: do not use if allergic reaction occurs or if patient has known allergies to device components.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Follow directions to remove if experiencing swelling, numbness, discomfort, pain, discoloration, or abnormal appearance.Note: if experiencing problems with use of the device, please consult your healthcare professional for assistance.Directions to apply: verify correct size prior to use.Trim pubic hair if necessary.Wash penis with mild soap and warm water.Dry thoroughly.Wear time may be reduced if skin is not dry or cream/oil is used.Open package at perforation.Remove catheter from plastic insert, if present.Place rolled end over the end of the penis, leaving a small space between the end of the penis and the cone of the catheter.Unroll the catheter over penis.Gently squeeze the catheter to properly seal adhesive to the skin.If possible, avoid leaving a rolled ¿collar¿ around the base of the penis.Important: wear time may be reduced if adhesive is not properly sealed to the skin.Connect the catheter to a drainage system.Make sure to check that connections are secure before use.Directions to remove: ensure drainage bag is empty.Disconnect catheter from the drainage system.Gently roll the catheter forward and off the penis.If necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.Disposal: after removal, dispose by placing the used catheter into a waste container.
 
Event or Problem Description
It was reported that the patient needed help with how to use external catheters.There was no malfunction with the device.The patient would contact liberator or doctor for further instruction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX SELF-ADHERING MALE EXTERNAL CATHETER
Common Device Name
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key12367180
Report Number1018233-2021-05183
Device Sequence Number19826945
Product Code NNX
UDI-Device Identifier00801741070730
UDI-Public(01)00801741070730
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number33301
Device Catalogue Number33301
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/30/2021
Initial Report FDA Received Date08/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-