• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EXPRT HIP; EXPRT REVISION HIP, LATERAL OFFSET PROXIMAL BODY, 75MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EXPRT HIP; EXPRT REVISION HIP, LATERAL OFFSET PROXIMAL BODY, 75MM Back to Search Results
Model Number 495-01-075
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/04/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to this document as soon as it becomes available.
 
Event or Problem Description
Revision surgery - patient had a stryker dual mobility acetabular cup in and on a expert stem.Surgeon went in and replaced the proximal body and the ceramic head because the patient was not stable.
 
Additional Manufacturer Narrative
The reason for this revision surgery was reported as instability.The actual length of in-vivo for the item listed is unknown as the original surgery date was not provided or could be established.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to instability.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities or trauma.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Additional Manufacturer Narrative
1644408-2021-00895 was reassessed and determined to be non-reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT HIP
Common Device Name
EXPRT REVISION HIP, LATERAL OFFSET PROXIMAL BODY, 75MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
Manufacturer Contact
kiersten soderman
9800 metric blvd.
austin, TX 78758-5445
MDR Report Key12367348
Report Number1644408-2021-00895
Device Sequence Number4115533
Product Code KWZ
UDI-Device Identifier00190446153856
UDI-Public(01)00190446153856
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161610
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/21/2026
Device Model Number495-01-075
Device Catalogue Number495-01-075
Device Lot Number547U1097
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/04/2021
Supplement Date Received by Manufacturer08/25/2021
10/05/2021
Initial Report FDA Received Date08/25/2021
Supplement Report FDA Received Date09/02/2021
12/29/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-