Model Number 71173547
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during an internal fixation surgery, a large screwdriver handle broke during use.Instruments were outside the patient.Surgery was completed with a backup device instead, without any delay.No harm to the patient reported.
 
Additional Manufacturer Narrative
G3, h2, h3, and h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.G2: report source update.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE SCREWDRIVER HANDLE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12367465
Report Number1020279-2021-06502
Device Sequence Number11866786
Product Code HSB
UDI-Device Identifier03596010519580
UDI-Public03596010519580
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71173547
Device Catalogue Number71173547
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/08/2021
Supplement Date Received by Manufacturer09/21/2021
Initial Report FDA Received Date08/25/2021
Supplement Report FDA Received Date09/22/2021
Patient Sequence Number1
-
-