• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RIVERPOINT MEDICAL LLC VELSORB FAST; SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RIVERPOINT MEDICAL LLC VELSORB FAST; SUTURE Back to Search Results
Model Number SV-9913
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Velosorb product were labeled as non-sterile and intended for shipment to the contract sterilizer but were inadvertently shipped to the distributor.Riverpoint placed an immediate hold on manufacturing and distribution of the velosorb product and a device recall (z-0010-2021).Corrective actions were taken by riverpoint to prevent incorrect shipping of non-sterile product.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
 
Event or Problem Description
Velosorb products were labeled as non-sterile and intended for shipment to the contract sterilizer but were inadvertently shipped to the distributor.Products were further distributed to customers.Riverpoint placed an immediate hold on manufacturing and distribution of the velosorb product and device recall z-0010-2021 was initiated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VELSORB FAST
Common Device Name
SUTURE
Manufacturer (Section D)
RIVERPOINT MEDICAL LLC
825 ne 25th ave
portland OR 97232
MDR Report Key12367696
Report Number3006981798-2021-00011
Device Sequence Number3793972
Product Code GAM
UDI-Device Identifier20884521153278
UDI-Public20884521153278
Combination Product (Y/N)N
PMA/510(K) Number
K120556
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 08/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date04/30/2023
Device Model NumberSV-9913
Device Lot Number20050709FX
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2020
Initial Report FDA Received Date08/25/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0010-2021
Patient Sequence Number1
-
-