MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ANTHEM RF CRT-P; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3210
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Arrhythmia (1721)
Date of Event 07/10/2021
Type of Reportable Event Death
Event or Problem Description
Related manufacturer reference number: 2017865-2021-28993.Related manufacturer reference number: 2017865-2021-28994.It was reported that the patient has deceased.There is no known allegation from a health care professional that suggests the death was device related.The cause of death was cardiac arrest.No additional information was reported.
 
Additional Manufacturer Narrative
The device was received from the field with battery voltage above elective replacement indicator (eri) level.Analysis of device image and longevity assessment testing indicated that the device was within normal range of operation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHEM RF CRT-P
Common Device Name
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
MDR Report Key12368109
Report Number2017865-2021-28989
Device Sequence Number14536272
Product Code NKE
UDI-Device Identifier05414734503785
UDI-Public05414734503785
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
P030035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2013
Device Model NumberPM3210
Device Catalogue NumberPM3210
Device Lot Number3653154
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 08/05/2021
Supplement Date Received by Manufacturer09/15/2021
Initial Report FDA Received Date08/25/2021
Supplement Report FDA Received Date10/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TENDRIL STS; TENDRIL STS; TENDRIL STS; TENDRIL STS
Outcome Attributed to Adverse Event Death;
-
-