| Model Number |
SV-2239 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Velosorb product were labeled as non-sterile and intended for shipment to the contract sterilizer but were inadvertently shipped to the distributor.Riverpoint placed an immediate hold on manufacturing and distribution of the velosorb product and a device recall (z-0021-2021).Corrective actions were taken by riverpoint to prevent incorrect shipping of non-sterile product.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
|
| |
|
Event or Problem Description
|
|
Velosorb products were labeled as non-sterile and intended for shipment to the contract sterilizer but were inadvertently shipped to the distributor.Products were further distributed to customers.Riverpoint placed an immediate hold on manufacturing and distribution of the velosorb product and device recall z-0021-2021 was initiated.
|
| |
|
Search Alerts/Recalls
|