U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX STERILIZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STERIS MEXICO, S. DE R.L. DE C.V. V-PRO MAX STERILIZER Back to Search Results
Model Number V-PRO maX
Medical Device Problem Codes Fluid/Blood Leak (1250); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A steris service technician arrived onsite to inspect the sterilizer and found it to be not operational.The damage was contained within the footprint of the unit.Prior to the reported event, the unit had alarmed "chamber overtemp".The user facility was provided with a loaner unit.Steris has requested the unit subject of the event be returned for evaluation; we are pending release of the unit by the customer.A follow-up mdr will be submitted when additional information becomes available.
 
Event or Problem Description
The user facility reported that their v-pro max sterilizer caught fire.No report of injury.
 
Additional Manufacturer Narrative
The v-pro max sterilizer subject of the event was returned to steris for evaluation.The evaluation found that the tee underneath the reservoir was leaking sterilant.The sterilant leaked onto the sterilizer's components resulting in the reported event.The tee should be replaced annually in accordance with the v-pro max sterilizer preventive maintenance checklist.The unit was installed in 2018 and is not under steris service agreement for maintenance activities; the user facility is responsible for all maintenance activities.No additional issues have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-PRO MAX STERILIZER
Common Device Name
STERILIZER
Manufacturer (Section D)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX  67190
Manufacturer (Section G)
STERIS MEXICO, S. DE R.L. DE C.V.
avenida avante 790
parque industrial guadalupe
guadalupe, 67190
MX   67190
Manufacturer Contact
daniel davy
5960 heisley road
mentor, OH 44060
4403927453
MDR Report Key12369796
Report Number3005899764-2021-00047
Device Sequence Number4496981
Product Code MLR
UDI-Device Identifier00724995206987
UDI-Public00724995206987
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberV-PRO maX
Device Catalogue NumberVP30002101
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2021
Supplement Date Received by Manufacturer07/27/2021
07/27/2021
Initial Report FDA Received Date08/26/2021
Supplement Report FDA Received Date11/12/2021
07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-