| Model Number |
V-PRO maX |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A steris service technician arrived onsite to inspect the sterilizer and found it to be not operational.The damage was contained within the footprint of the unit.Prior to the reported event, the unit had alarmed "chamber overtemp".The user facility was provided with a loaner unit.Steris has requested the unit subject of the event be returned for evaluation; we are pending release of the unit by the customer.A follow-up mdr will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The user facility reported that their v-pro max sterilizer caught fire.No report of injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The v-pro max sterilizer subject of the event was returned to steris for evaluation.The evaluation found that the tee underneath the reservoir was leaking sterilant.The sterilant leaked onto the sterilizer's components resulting in the reported event.The tee should be replaced annually in accordance with the v-pro max sterilizer preventive maintenance checklist.The unit was installed in 2018 and is not under steris service agreement for maintenance activities; the user facility is responsible for all maintenance activities.No additional issues have been reported.
|
| |
|