• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 2/150/150; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND PASSEO-18 2/150/150; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 366129
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/16/2021
Type of Reportable Event Malfunction
Event or Problem Description
The physician intended to treat a severely calcified lesion (90 percent stenosis degree, vessel reference diameter 6 mm, lesion length 180 mm) in a mildly tortuous bifurcation of the left anterior tibial artery.A passeo-18 balloon catheter was used for the procedure.Initially, it was reported that the balloon of the complaint instrument burst and ruptured in the vessel.After additional correspondence with the local staff it was clarified that the complaint instrument was successfully inflated three times.After balloon deflation, the instrument was withdrawn during which the balloon tore off the catheter.Since the balloon could not be recovered, the patient underwent surgery.The balloon was successfully removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PASSEO-18 2/150/150
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key12370304
Report Number1028232-2021-04800
Device Sequence Number4015035
Product Code LIT
UDI-Device Identifier07640130414716
UDI-Public07640130414716
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072765
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model Number366129
Device Catalogue NumberSEE MODEL NO.
Device Lot Number11202139
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2021
Initial Report FDA Received Date08/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-