Model Number 0998-00-3013-67
Device Problems Failure to Power Up (1476); Failure to Pump (1502)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2021
Event Type  malfunction  
Manufacturer Narrative
Device not accessible for testing: additional information is being requested with regard to the repair and status of the iabp.A supplemental report will be submitted if this information is provided to us.
 
Event Description
It was reported that the cs100 intra-aortic balloon pump failed to power up.It is unknown if there was patient involvement.There was no adverse event reported.
 
Manufacturer Narrative
Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to evaluate the iabp.To fix the issue, the fse replaced the helium outlet fitting because it was observed to be broken.All functional and safety checks were performed to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the cs100 intra-aortic balloon pump does not start.It was later clarified that the iabp unit was unable to be started as it would not fill with helium.It is unknown under which circumstances this event occurred; however there was no patient involvement, and no adverse event was reported.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, g8, h2, h4, h6 (type of investigation, investigation findings, investigation conclusions), h10, h11.Corrected field: g1 (contact person ¿ mfg site).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS100 INTRA-AORTIC BALLOON PUMP, ITALIAN, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12373234
MDR Text Key268329613
Report Number2249723-2021-01916
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107486
UDI-Public10607567107486
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3013-67
Device Catalogue Number0998-00-3013-67
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received07/29/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/30/2005
Is the Device Single Use? No
Type of Device Usage A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-