MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEO GASTROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; LINEAR ULTRASOUND VIDEO GASTROSCOPE Back to Search Results
Model Number EG-3270UK
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If additional information becomes available, a supplemental report will be filed with the new information.
 
Event or Problem Description
Pentax medical was made aware of an event that occurred in the united states.The information provided indicated that the suction channel was blocked involving a pentax medical ultrasound video gastroscope model eg-3270uk, serial number (b)(4).The customer stated there was an internal leak inside the channel.Good faith efforts(gfe) emails have been sent.No additional information has been provided.The customer owned endoscope was received by pentax medical for evaluation on 02-aug-2021.The endoscope was inspected by pentax medical service under service order (b)(4) and during quality inspection the technician confirmed the customers experience as the "operation channel- primary slice by accessory" was identified and also documented the following inspection findings: leak at biopsy channel (large/ primary) distal side, failed wet leak test, failed dry leak test, suction tube mild resistance, ultrasound warning label cut, suction function not performed unit compromised, operation channel- primary slice by accessory, fluid invasion not observed in pve connector, fluid invasion not observed in control body.The device underwent repairs including the following components: o-rings and seals, bending rubber, operation channel, suction channel lg, warning label.Model eg-327uk, serial number (b)(4), has been routinely serviced at a pentax facility since the device was put into service.The endoscope was approved by final qc on 19-aug-2021 and delivered to the customer under delivery order (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Common Device Name
LINEAR ULTRASOUND VIDEO GASTROSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0012
JA 
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller (temporary)
3 paragon drive
montvale, NJ 07645
8004315880
MDR Report Key12373950
Report Number9610877-2021-00663
Device Sequence Number3935322
Product Code ODG
UDI-Device Identifier04961333233793
UDI-Public04961333233793
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K182004
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEG-3270UK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/27/2021
Supplement Date Received by Manufacturer07/27/2021
Initial Report FDA Received Date08/26/2021
Supplement Report FDA Received Date10/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/22/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-