• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; LIGATION / VESSEL CLIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG LIGATURE CLIP 12 MAG.= 144 PCS.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL572T
Medical Device Problem Code Unintended Ejection (1234)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/29/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event or Problem Description
It was reported that pl572t was used during laparoscopic procedure performed on (b)(6) 2021.According to the complaint description, during the second shot, the clip was clogged, the clip magazine body fell off from pl604r, and the magazine fell into the patient's abdominal cavity.An additional medical intervention was necessary.Additional information was not provided.Additional patient information is not available.The adverse event is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Correction / additional information: b5: product name.H6: codes.Investigation results: visual investigation: the investigation was carried out visually and microscopically with the digital-camera and broken off noses, broken off latches and clip jams were found.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 3(5) x probability of occurrence 3(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a psc 2021-113 has been initiated.Any action regarding capa will be addressed with this case.
 
Event or Problem Description
Correction: added the "product name".It was reported that there was an issue with pl572t - ligature clip 12 mag.= 144 pcs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGATURE CLIP 12 MAG.= 144 PCS.
Common Device Name
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key12373989
Report Number9610612-2021-00584
Device Sequence Number19824539
Product Code FZP
UDI-Device Identifier04038653341856
UDI-Public4038653341856
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K081031
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2025
Device Model NumberPL572T
Device Catalogue NumberPL572T
Device Lot Number52645768
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2021
Supplement Date Received by Manufacturer04/04/2022
Initial Report FDA Received Date08/26/2021
Supplement Report FDA Received Date04/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-