ETHICON ENDO-SURGERY, LLC. ENSEAL G2 STRAIGHT JAW 45CM; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSLG2S45
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: r93p1k, r93g48.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This is a analysis for a set of images submitted to ethicon endo surgery for evaluation.The image provided by the customer is that of the outer label from a shipping box, the lot can be observed as r93p1k.The image provided by the customer is that of the label from the secondary package.The lot can be observed as r93g48.Based on the information currently available, a product issue was identified during the investigation of the sample received.This product issue will be addressed through ethicon endo surgery quality system.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.
 
Event or Problem Description
It was reported that during an unknown procedure a lot mismatch between the outer box which reports r93p1k (exp.31/5/23) and the inner box which reports r93g48 (exp.30/4/23).
 
Additional Manufacturer Narrative
(b)(4).Date sent: 10/6/2021.
 
Additional Manufacturer Narrative
(b)(4), date sent: 11/9/2021 additional information was requested and the following was obtained: the distribution center this complaint was reported from is a jnj distribution center.The product was still under jnj control investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection was conducted on the returned device.The nslg2s45 device a was returned inside its original packaging.Upon visual inspection, it was noted that the outer label from a shipping box, the lot can be observed as r93p1k, and the label from the internal box and the tyvek a lot can be observed as r93g48.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon endo surgery quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Further investigation on the complaint a manufacturing issue was found at (b)(4), an nr was initiated to documented it.Upon review of the information provided, it was concluded that this event does not meet the fda defined criteria for a reportable event and is being considered not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL G2 STRAIGHT JAW 45CM
Common Device Name
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
6107428552
MDR Report Key12374577
Report Number3005075853-2021-05009
Device Sequence Number17160573
Product Code GEI
UDI-Device Identifier10705036014195
UDI-Public10705036014195
Combination Product (Y/N)N
PMA/510(K) Number
K112033
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model NumberNSLG2S45
Device Catalogue NumberNSLG2S45
Device Lot NumberR93P1K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2021
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/02/2021
Supplement Date Received by Manufacturer09/20/2021
10/13/2021
Initial Report FDA Received Date08/26/2021
Supplement Report FDA Received Date10/06/2021
11/09/2021
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/25/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GENERATOR
Patient SexUnknown
-
-