SMITH & NEPHEW, INC. PLATE CUTTING FORCEPS; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71170047 |
Device Problems
Break (1069); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/11/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that a plate cutting forceps is stiff and does not re-open after use, also the cutting blades are chipped.As this was noticed upon inspection, there was not patient involvement.
|
|
Manufacturer Narrative
|
H3, h6: the device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures for the listed batch; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|
Search Alerts/Recalls
|
|
|