• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. PLATE CUTTING FORCEPS; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. PLATE CUTTING FORCEPS; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71170047
Device Problems Break (1069); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2021
Event Type  malfunction  
Event Description
It was reported that a plate cutting forceps is stiff and does not re-open after use, also the cutting blades are chipped.As this was noticed upon inspection, there was not patient involvement.
 
Manufacturer Narrative
H3, h6: the device, intended for use in treatment, was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A complaint history review found related failures for the listed batch; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATE CUTTING FORCEPS
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12374645
MDR Text Key268472906
Report Number1020279-2021-06572
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010180803
UDI-Public03596010180803
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 10/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71170047
Device Catalogue Number71170047
Device Lot Number15DTDK016
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-