• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL QUATTRODE LEAD, 3/4MM, 90 CM; PERCUTANEOUS LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL QUATTRODE LEAD, 3/4MM, 90 CM; PERCUTANEOUS LEAD Back to Search Results
Model Number 3149
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 06/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event is estimated.
 
Event or Problem Description
Related manufacturer reference numbers: 1627487-2021-16611, 1627487-2021-16613, 1627487-2021-16614.It was reported that the patient experienced ineffective stimulation.Diagnostics revealed high impedances on 8 contacts for two leads.Reprogramming was attempted, but was unable to restore therapy.In turn, surgical intervention may take place at a later date to address the issue.It is unknown which two leads have impedance issues, therefore all leads are being reported.
 
Additional Manufacturer Narrative
During processing of this complaint, attempts were made to obtain complete patient information.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Additional information received indicates the lead was replaced to address the issue.Patient had effective therapy post operatively.
 
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRODE LEAD, 3/4MM, 90 CM
Common Device Name
PERCUTANEOUS LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key12374827
Report Number1627487-2021-16612
Device Sequence Number16442629
Product Code LGW
UDI-Device Identifier05414734401593
UDI-Public05414734401593
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2014
Device Model Number3149
Device Catalogue Number3149
Device Lot Number3629523
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2021
Supplement Date Received by Manufacturer11/11/2021
05/03/2022
06/04/2022
Initial Report FDA Received Date08/26/2021
Supplement Report FDA Received Date11/17/2021
05/27/2022
06/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/29/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PERCUTANEOUS LEAD (X3)
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-