| Model Number |
72203852 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Unspecified Tissue Injury (4559); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an arthroscopy procedure, suture fix anchors pulled out, a new bone hole was drilled.The procedure was successfully completed without significant delay using a back-up device.No other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an arthroscopy procedure, the suture fix anchors pulled out.A new bone hole was drilled.The procedure was successfully completed without significant delay using a back-up device.No other complications were reported.The patient current status is good.
|
| |
|
Search Alerts/Recalls
|