MAUDE Adverse Event Report: PENUMBRA, INC. POD COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. POD COIL; HCG, KRD Back to Search Results
Model Number RBYPOD5-A
Medical Device Problem Codes Physical Resistance/Sticking (4012); Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2021-01909, 3005168196-2021-01911.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the splenic artery using a ruby coil detachment handle (handle), pod coils, a non-penumbra microcatheter, a non-penumbra catheter, and a non-penumbra guide catheter.During the procedure, the physician advanced a pod coil into the target location and attempted to detach it using a handle; however, the pod coil failed to detach after multiple attempts.It was reported that the coil alignment zone separated, but the pod coil did not detach.Therefore, the physician pulled the pusher assembly of the pod coil with the handle to detach it.It was also reported that the physician experienced resistance while inserting the proximal end of the pusher assembly into the handle.Next, the physician advanced another pod coil to the target location and attempted to detach it using the handle; however, the same issue occurred.Therefore, the physician also pulled the pusher assembly of the pod coil with the handle to detach it.The procedure was completed using another handle and eight additional coils.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that the following section was inadvertently missed on the initial mfr report and is being updated on this follow-up # 02 mfr report:3005168196-2021-01910: section d box 6.If implanted, give date.This report is associated with mfr report number: 3005168196-2021-01909; 3005168196-2021-01911.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key12375952
Report Number3005168196-2021-01910
Device Sequence Number4116228
Product Code HCG
UDI-Device Identifier00814548019604
UDI-Public00814548019604
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K173614
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/27/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRBYPOD5-A
Device Catalogue NumberRBYPOD5
Device Lot NumberF96238
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/02/2021
Supplement Date Received by Manufacturer01/14/2005
Initial Report FDA Received Date08/27/2021
Supplement Report FDA Received Date12/27/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-