• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. ANALYZER SURGICAL CABLE; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. ANALYZER SURGICAL CABLE; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2292
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/31/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that when performing detection tests for implanted leads using the analyzer surgical cable with the programmer it was observed that when connecting only the atrium pins (blue pins) in the lead, the programmer showed the detection value in the ventricle section and when connecting only the ventricle pins (white pins) in the lead, the programmer showed the detection value in the atrium section.It was noted that the analyzer surgical cable was switched out for a different cable which resolved the issue.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product analysis: analysis could not confirm the customer's comment that there was an issue with the analyzer surgical cable as the cable passed all visual inspection and bench testing with no anomalies observed.Visual inspection found no issues and all continuity tests were ok.Correct resistance was found between pins 3 and 5 and no intermittent or shorted connections found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANALYZER SURGICAL CABLE
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key12376599
Report Number2182208-2021-03382
Device Sequence Number17301488
Product Code KRG
UDI-Device Identifier00643169703247
UDI-Public00643169703247
Combination Product (Y/N)N
Initial Reporter CountryTS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2292
Device Catalogue Number2292
Device Lot Number752470001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2021
Initial Date Received by Manufacturer 08/02/2021
Supplement Date Received by Manufacturer09/17/2021
Initial Report FDA Received Date08/27/2021
Supplement Report FDA Received Date10/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2090 PROGRAMMER.; 2090 PROGRAMMER
-
-