• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC NEO METER PRECISION FREESTYLE; BLOOD GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC NEO METER PRECISION FREESTYLE; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71741-80
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An extended investigation has been conducted, which involved a manufacturing review (including 5 years of device history records (dhrs)), previous complaint and corrective and preventive action (capa) investigations conducted for blank screen issues, process failure mode effects analyses (pfmeas), design controls, and design failure mode effects analyses (dfmeas), risk management reports, risk evaluations, and label copy.The investigation did not identify any indication that the product did not meet specification.If the product is returned, an investigation will be performed.All pertinent information available to abbott diabetes care has been submitted.
 
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEO METER PRECISION FREESTYLE
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key12377032
Report Number2954323-2021-82021
Device Sequence Number4067320
Product Code NBW
UDI-Device Identifier00357599000301
UDI-Public00357599000301
Combination Product (Y/N)N
PMA/510(K) Number
K140371
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number71741-80
Device Catalogue Number71741
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2021
Initial Report FDA Received Date08/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age43 YR
Patient Weight68
-
-