DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-3023-53 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/02/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device: the getinge field service engineer (fse) that encountered the issue, replaced the power cord to correct the problem, and completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted when our investigation is completed.The initial reporter named in is a getinge employee who has different contact details from that of the event site.A contact person at the event site is biomed (b)(6).
|
|
Event Description
|
It was reported that during a preventive maintenance (pm) performed by a getinge field service engineer (fse) on the cs300 intra-aortic balloon pump (iabp), the iabp failed the ground continuity test (electrical ground resistance test).There was no patient involvement, and no adverse event reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, g8, h2, h6(investigation type, investigation findings & investigation conclusions), h10, h11 corrected fields: d1, d4(udi#), e4, g1(contact person).
|
|
Event Description
|
N/a.
|
|