Model Number 0998-00-3023-53
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/02/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device: the getinge field service engineer (fse) that encountered the issue, replaced the power cord to correct the problem, and completed the pm with full calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted when our investigation is completed.The initial reporter named in is a getinge employee who has different contact details from that of the event site.A contact person at the event site is biomed (b)(6).
 
Event Description
It was reported that during a preventive maintenance (pm) performed by a getinge field service engineer (fse) on the cs300 intra-aortic balloon pump (iabp), the iabp failed the ground continuity test (electrical ground resistance test).There was no patient involvement, and no adverse event reported.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, g8, h2, h6(investigation type, investigation findings & investigation conclusions), h10, h11 corrected fields: d1, d4(udi#), e4, g1(contact person).
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12377479
MDR Text Key268470146
Report Number2249723-2021-01921
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/15/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/30/2008
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-