| Model Number |
627233S |
| Medical Device Problem Codes |
Fracture (1260); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned.If additional information becomes available, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported the holes for k-wire/suture on a stryker proximal humerus plate have a fiber wire torn/snapped.The device was implanted on tubercle minus to fix/fixate on the proximal humerus.The issue was observed during a revision surgery performed on (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported the holes for k-wire/suture on a stryker proximal humerus plate have a fiber wire torn/snapped.The device was implanted on tubercle minus to fix/fixate on the proximal humerus.The issue was observed during a revision surgery performed on (b)(6) 2021.The revision surgery was performed because during the primary operation the reposition of the fragments needed to be done again.
|
| |
|