ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 03P6822
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/20/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There was no further patient information provided by the customer.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed falsely elevated magnesium results on architect c4000 processing module for multiple patients.The few results examples provided were: on (b)(6) 2021 patient 1: initial=3.9 mmol/l /repeated=0.71 mmol/l.Patient 2: initial=3.2 mmol/l /repeated=0.67 mmol/l there was no reported impact to patient management.
 
Additional Manufacturer Narrative
Cross reference mdr for likely cause changed; magnesium to architect c4000 processing module, list 02p24-40, 3016438761-2021-00323-00.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
MDR Report Key12378047
Report Number3002809144-2021-00527
Device Sequence Number3938329
Product Code JGJ
UDI-Device Identifier00380740161620
UDI-Public00380740161620
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K181748
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03P6822
Device Catalogue Number03P68-22
Device Lot Number65953UN21
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/23/2021
Supplement Date Received by Manufacturer08/30/2021
Initial Report FDA Received Date08/27/2021
Supplement Report FDA Received Date09/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C4000 INTGR, (B)(4); ARC C4000 INTGR, (B)(4); ARC C4000 INTGR, (B)(4)
-
-