| Model Number |
CAREAWARE 6.5.0 CAREAWARE IBUS.2021W16 THROUGH CAREAWARE IBUS.2021W30 |
| Medical Device Problem Code |
Computer Software Problem (1112)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Cerner distributed a flash notification on august 26, 2021 to all potentially impacted client sites.The software notification includes a description of the issue and a software modification has been developed to address the issue for all sites that could be potentially impacted.Cerner corporation considers this issue to be resolved and no further narrative is required for follow-up.
|
| |
|
Event or Problem Description
|
|
This report documents information related to an issue identified with functionality included in cerner's careaware ibus®.This issue involves cerner's careaware ibus® and affects users that assign role based alerting in careaware® event management, which is used as a secondary mechanism to route event or alarm information to mobile phones or other computing or communications endpoints.If a personnel record is saved in cerner millennium, that personnel record can fail to set an organization identifier in careaware event management due to a defect in careaware ibus.If alerting is built based on roles, the clinician may not be notified appropriately of a secondary alert through careaware event management, which could lead to a delay in patient care.Cerner has not received communication on any adverse patient events as a result of this issue.
|
| |
|