| Model Number |
304310 |
| Medical Device Problem Codes |
Failure to Osseointegrate (1863); Osseointegration Problem (3003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/29/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No evidence was found with the implant itself after evaluation.Customer did state they were unable to achieve primary stability.
|
| |
|
Event or Problem Description
|
|
Asymptomatic and inadequate bone quality/bone quantity were reported.Bone quality at the time of implant failure type iii.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability and osseointegration was not achieved.Augmentation procedure was not performed at the time of surgery.Patient has arthritis.
|
| |
|