• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012453-20
Medical Device Problem Codes Inflation Problem (1310); Stretched (1601)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual and functional inspections were performed on the returned device.The reported inflation issue was not confirmed.However, the analysis identified the inner and outer member was stretched and bunched distal to the mid lap seal.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported inflation issue.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of this device.
 
Event or Problem Description
It was reported that the procedure was to treat a de novo, moderately calcified, mildly tortuous lesion in the left anterior descending (lad) artery.The 4.0x20 mm nc trek balloon dilatation catheter (bdc) was able to hold pressure but could only partially inflate inside the previously implanted stent.There was no difficulty in removing the stylet or protective sheath.Another nc trek bdc was used to successfully complete the procedure.There were no reported adverse patient effects and there was no clinically significant delay in the procedure.Returned device analysis identified a stretched inner/outer member.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12378821
Report Number2024168-2021-07597
Device Sequence Number3938549
Product Code LOX
UDI-Device Identifier08717648152085
UDI-Public08717648152085
Combination Product (Y/N)N
Initial Reporter CountryEG
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2022
Device Model Number1012453-20
Device Catalogue Number1012453-20
Device Lot Number90226G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/06/2021
Initial Report FDA Received Date08/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
Patient Weight78
-
-