• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARL ZEISS MEDITEC AG (JENA) VISUMAX; FEMTOSECOND LASER SYSTEM FOR REFRACTIVE CORRECTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARL ZEISS MEDITEC AG (JENA) VISUMAX; FEMTOSECOND LASER SYSTEM FOR REFRACTIVE CORRECTION Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Date of Event 04/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
According to the device evaluation results the visumax worked within specification.The manufacturer determined that the root cause of the event is user error (decentration).Checking of the pre-op and the post-op topography data shows that the ablation was clearly shifted to the right (os).Review of the treatment video confirmed that there is a decentration.The decentered smile explains the visual outcome.Proper preparation of the patient and aspects that require additional attention as well as further safety and warning instructions are described in the user manual.
 
Event or Problem Description
A health care professional (hcp) reported that a patient who went through a smile procedure lost 3 lines of the best corrected visual acuity (bcva).The patient's one eye is 20/20 but the other eye's bcva is 20/50.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISUMAX
Common Device Name
FEMTOSECOND LASER SYSTEM FOR REFRACTIVE CORRECTION
Manufacturer (Section D)
CARL ZEISS MEDITEC AG (JENA)
carl zeiss promenade 10
jena, thueringia 7745
GM  7745
MDR Report Key12380094
Report Number9615030-2021-00017
Device Sequence Number14437045
Product Code OTL
UDI-Device Identifier04049471095036
UDI-Public(01)04049471095036(11)161101
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P150040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number000000-1345-518
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/29/2021
Initial Report FDA Received Date08/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age25 YR
-
-