| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Visual Impairment (2138)
|
| Date of Event |
04/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
According to the device evaluation results the visumax worked within specification.The manufacturer determined that the root cause of the event is user error (decentration).Checking of the pre-op and the post-op topography data shows that the ablation was clearly shifted to the right (os).Review of the treatment video confirmed that there is a decentration.The decentered smile explains the visual outcome.Proper preparation of the patient and aspects that require additional attention as well as further safety and warning instructions are described in the user manual.
|
| |
|
Event or Problem Description
|
|
A health care professional (hcp) reported that a patient who went through a smile procedure lost 3 lines of the best corrected visual acuity (bcva).The patient's one eye is 20/20 but the other eye's bcva is 20/50.
|
| |
|