| Model Number |
STRIP, LEADER TT MEDIMEDI 50CT |
| Medical Device Problem Codes |
High Test Results (2457); Low Test Results (2458)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
08/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal report reference number: (b)(4).Meter and test strips were not returned for evaluation.Note: manufacturer contacted customer in a follow-up call on 13-aug-2021 to ensure the replacement products resolved the initial concern - able to establish contact with customer who stated replacement products resolved initial concern.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high and low blood glucose test results.The customer is concerned with test results from results obtained of 258, 173, 73, 182 and 191mg/dl.The customer¿s expected fasting blood glucose test result range is 80-160mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call, a back to back blood test was not performed by the customer.The product is not stored according to specification (kitchen).The test strip lot manufacturer¿s expiration date is 02/22/2023 and open vial date was not disclosed.The customer did not have another vial of test strips that had been stored and handled correctly.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
Sections with additional information as of 02-nov-2021: h6: updated fda¿s type, findings and conclusions codes.H10: meter was not returned for evaluation.Test strips were not returned for evaluation.Retention testing was performed using test strips from the same lot.Retention strip lot tested within specifications.Most likely underlying root cause: mlc-020: user's test strip had poor storage.
|
| |
|
Search Alerts/Recalls
|