• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CONQUEST 40; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CONQUEST 40; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQF75104
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/30/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly had a pin hole rupture.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported failure as no objective evidence was provided for review.A definitive root cause for the alleged pinhole balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that during an angioplasty procedure, the pta balloon allegedly had a pin hole rupture.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST 40
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key12381189
Report Number2020394-2021-01521
Device Sequence Number4115622
Product Code DQY
UDI-Device Identifier00801741060632
UDI-Public(01)00801741060632
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K120660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQF75104
Device Catalogue NumberCQF75104
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Returned to Manufacturer08/25/2021
Initial Date Received by Manufacturer 08/02/2021
Supplement Date Received by Manufacturer09/21/2021
Initial Report FDA Received Date08/28/2021
Supplement Report FDA Received Date09/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-