BOSTON SCIENTIFIC CORPORATION ENCORE 26 ADVANTAGE KIT; SYRINGE, BALLOON INFLATION Back to Search Results
Model Number 2760
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/05/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was contaminated.An encore 26 advantage kit was selected for a percutaneous coronary intervention (pci).During the procedure, the device was used to inflate the percutaneous transluminal coronary angioplasty balloons.During post dilatation, small metallic particles were found after the chamber was filled.Contrast and heparinized saline was added.The source of the particles is unknown.The procedure was completed with another device.The patient's status was stable.
 
Event or Problem Description
It was reported that the device was contaminated.An encore 26 advantage kit was selected for a percutaneous coronary intervention (pci).During the procedure, the device was used to inflate the percutaneous transluminal coronary angioplasty balloons.During post dilatation, small metallic particles were found after the chamber was filled.Contrast and heparinized saline was added.The source of the particles is unknown.The procedure was completed with another device.The patient's status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the unit was returned in a generic plastic bag with batch 27392658 printed on it.The device was returned with an amount of saline solution within the barrel of the device; however, after a complete visual and microscopic inspection, no contaminant agents or foreign matters were observed within the encore device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE 26 ADVANTAGE KIT
Common Device Name
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key12384210
Report Number2134265-2021-10892
Device Sequence Number11924317
Product Code MAV
UDI-Device Identifier08714729180005
UDI-Public08714729180005
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K140745
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/27/2023
Device Model Number2760
Device Catalogue Number2760
Device Lot Number0027392658
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2021
Initial Date Received by Manufacturer 08/09/2021
Supplement Date Received by Manufacturer09/07/2021
Initial Report FDA Received Date08/30/2021
Supplement Report FDA Received Date09/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-