• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 4 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US TRI-LOCK BPS SZ 4 STD OFFSET; TRILOCK HIP STEM : HIP FEMORAL STEM Back to Search Results
Model Number 1012-04-040
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526)
Date of Event 08/10/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient presented with increasing hip pain and instability.Trilock stem is noted to have subsided at some point in time.Stem was well fixed and difficult to remove.40mm +5 metal head was also explained (product and lot numbers could not be discerned).Cup and liner reminder insitu.Doi: (b)(6) 2008, dor: (b)(6) 2021, right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : a device history record (mre) review, was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-LOCK BPS SZ 4 STD OFFSET
Common Device Name
TRILOCK HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12384224
Report Number1818910-2021-18880
Device Sequence Number3650610
Product Code KWA
UDI-Device Identifier10603295000754
UDI-Public10603295000754
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012-04-040
Device Catalogue Number101204040
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 08/10/2021
Supplement Date Received by Manufacturer09/21/2021
Initial Report FDA Received Date08/30/2021
Supplement Report FDA Received Date09/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TRI-LOCK BPS SZ 4 STD OFFSET; UNK 40+5 METAL HEAD; TRI-LOCK BPS SZ 4 STD OFFSET; UNK 40+5 METAL HEAD
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient Weight95
-
-