Model Number 0998-00-0800-53
Device Problems Noise, Audible (3273); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/04/2021
Event Type  malfunction  
Manufacturer Narrative
Testing of actual/suspected device : a getinge field service engineer (fse) evaluated the iabp unit and found codes 58 and 124 in the fault logs.When the fse opened the pneumatic module assembly (pim), he found it to be full of liquid.To fix the issue, the fse replaced the pim, and performed a full preventive maintenance with all calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted when our investigation is completed.The full initial reporter name in is (b)(6).
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) shut down with a loud screaming noise.No patient harm, serious injury or adverse event was reported.
 
Manufacturer Narrative
Updated fields: b4, g3, g6, g7, h2, h4, h6, h10.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYP B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key12384513
MDR Text Key268692389
Report Number2249723-2021-01948
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/16/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-