DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYP B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problems
Noise, Audible (3273); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Testing of actual/suspected device : a getinge field service engineer (fse) evaluated the iabp unit and found codes 58 and 124 in the fault logs.When the fse opened the pneumatic module assembly (pim), he found it to be full of liquid.To fix the issue, the fse replaced the pim, and performed a full preventive maintenance with all calibration, functional and safety checks to meet factory specifications.Unit passed all calibration, functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted when our investigation is completed.The full initial reporter name in is (b)(6).
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) shut down with a loud screaming noise.No patient harm, serious injury or adverse event was reported.
|
|
Manufacturer Narrative
|
Updated fields: b4, g3, g6, g7, h2, h4, h6, h10.
|
|
Event Description
|
N/a.
|
|