CONMED UTICA ADJUSTABLE RETRACTOR CANNULA 8.0MM X 65MM; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Catalog Number C08-65
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported device is being returned to conmed for evaluation.A supplemental and final report will be filed following the completion of the device evaluation and complaint investigation.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Event or Problem Description
The sales representative reported on behalf of the customer that the device, c08-65, was being used on (b)(6) 2021 during a rotator cuff repair procedure and the ¿surgeon was inserting cannula into shoulder and 1 cannula partial distal flange snapped off.Opened second cannula and whole distal tip and flange broke off.All part of cannula removed safely from patient.¿ there was a two minute delay and the procedure was completed with an arthrex passport cannula.The arthrex passport cannula will be listed as a concomitant device.There was no patient injury or impact to the 75 year old, male patient weighing 98kg and no further intervention required.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Event or Problem Description
The sales representative reported on behalf of the customer that the device, c08-65, was being used on 8jul21 during a rotator cuff repair procedure and the ¿surgeon was inserting cannula into shoulder and 1 cannula partial distal flange snapped off.Opened second cannula and whole distal tip and flange broke off.All part of cannula removed safely from patient.¿ there was a two minute delay and the procedure was completed with an arthrex passport cannula.The arthrex passport cannula will be listed as a concomitant device.There was no patient injury or impact to the 75 year old, male patient weighing 98kg and no further intervention required.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
 
Additional Manufacturer Narrative
Customer event ¿cannula partial distal flange snapped off¿ was confirmed based on photographic evidence and device evaluation.Visual examination of the returned used product found both cannulas broken off device at the tip.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of two complaints, regarding two devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be 0.0002.Per the instructions for use, the user is advised the following: inspect cannula prior to use to ensure they are in good physical condition.To avoid damaging seals and excessive leakage, do not use with devices larger than the specified cannula diameter.Use carefully to ensure the cannula is not bent or collapsed when inserting devices.To avoid damage during use, do not use excessive force on cannula.This issue will continue to be monitored through the complaint system to assure patient safety.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADJUSTABLE RETRACTOR CANNULA 8.0MM X 65MM
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
CONMED UTICA
525 french rd
utica NY 13502
MDR Report Key12385541
Report Number1320894-2021-00329
Device Sequence Number20037486
Product Code NBH
UDI-Device Identifier00653405001631
UDI-Public(01)00653405001631(17)260404(10)202104051
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/04/2026
Device Catalogue NumberC08-65
Device Lot Number202104051
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2021
Initial Date Received by Manufacturer 07/08/2021
Supplement Date Received by Manufacturer09/13/2021
Initial Report FDA Received Date08/30/2021
Supplement Report FDA Received Date09/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTHREX PASSPORT CANNULA; ARTHREX PASSPORT CANNULA
Patient Age75 YR
Patient Weight98
-
-