• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL AGILIS NXT STEERABLE INTRODUCER DUAL-REACH 61 CM SHEATH, SMALL CURL DUAL-REACH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL AGILIS NXT STEERABLE INTRODUCER DUAL-REACH 61 CM SHEATH, SMALL CURL DUAL-REACH; INTRODUCER, CATHETER Back to Search Results
Model Number G408318
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
One 8.5f agilis introducer sheath was received for evaluation.The dilator was not returned.The sheath tip had been punctured, bent, scraped and was bulged and flared.The dilator was inserted into the sheath and a gap was noted at the dilator/sheath transition, consistent with the aforementioned damage to the sheath tip.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the sheath tip damage is consistent with damage during use.
 
Event Description
This report is to advise of an event observed during analysis confirming a puncture of the sheath tip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILIS NXT STEERABLE INTRODUCER DUAL-REACH 61 CM SHEATH, SMALL CURL DUAL-REACH
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key12385747
MDR Text Key268730904
Report Number3008452825-2021-00441
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734206112
UDI-Public05414734206112
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K061363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Model NumberG408318
Device Catalogue NumberG408318
Device Lot Number7599274
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/27/2021
Initial Date FDA Received08/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/24/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-