| Catalog Number |
0504860065 |
| Health Effect - Clinical Codes |
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip broke during procedure.The broken tip was successfully retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence.Alleged failure: tip broke off.Probable root cause: 1.Manufacturing/assembly error 2.Severe shipping conditions 3.Material/design error 4.Constant irrigation flow is not maintained leading to limited field of view 5.Low lubricity of insulation for fulgurating electrodes 6.Use error manufacture date is not known.H3 other text : 81.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip broke during procedure.The broken tip was successfully retrieved.
|
| |
|