• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97714
Device Problems Improper or Incorrect Procedure or Method (2017); Battery Problem (2885); Charging Problem (2892); Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/26/2021
Event Type  Injury  
Manufacturer Narrative
Date inaccurate, only the month and year are valid.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient regarding an implantable neurostimulator (ins) for non-malignant pain.The reason for call was the patient received the "reposition the antenna" screen on the recharger and could not charge their ins.They also received the "poor communication" screen on the programmer.The patient was redirected to their healthcare provider and manufacturer representative but the caller mentioned they contacted the representative on friday, and representative insisted they call in.Information was reviewed regarding a possible overdischarge, and that they would likely need to get the ins out of a possible overdischarged state with their healthcare provider and representative.A replacement recharger antenna was requested.Additional information received from the patient on 2021-aug-16.It was reported that replacement of the recharger antenna did not resolve the issue.They were "going to do replacement" when they had an appointment with their healthcare provider.No further complications reported.
 
Event Description
Additional information was received.It was reported, the cause of the charging/connection issue was, that it was overcharged.And the patient was told they needed a new one, as their current one was 4 years old.There was no surgery scheduled.As they were seeing their healthcare provider in october to see if it might be hip pain, for hip replacement.The patient was still waiting on their healthcare provider to decide what is going on.
 
Manufacturer Narrative
Concomitant medical products: product id: 37791, serial# unknown, product type: recharger.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Continuation of d10: product id 37791; serial# unknown; product type recharger.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from the mri technologist.They called, and were wondering if the patient could have an mri.They reiterated patient's ins battery was dead and they has discussed it with the manufacturer previously.It was reviewed that it is recommended mri mode were activated before having an mri scan.The patient would need to follow-up with the hcp/rep for troubleshooting.They reviewed that if mri mode cannot be activated hcp may be able to complete scan-type eligibility form, reviewed statement in mri labeling that says a depleted stimulator is considered off.The mri tech noted the patient had an mri before and it was confirmed with the controller that she was full-body conditional.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12387239
MDR Text Key271275857
Report Number3004209178-2021-13094
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169109483
UDI-Public00643169109483
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/28/2018
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received11/11/2021
Date Device Manufactured08/30/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
"SEE H10"; "SEE H10"
Patient Outcome(s) Required Intervention;
Patient Age77 YR
Patient SexFemale
Patient Weight73 KG
-
-