COOK INC GUNTHER TULIP JUGULAR VENA CAVA FILTER SET DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number G33017
Device Problem Structural Problem (2506)
Patient Problems Chest Pain (1776); Dyspnea (1816); Perforation of Vessels (2135)
Event Type  Injury  
Manufacturer Narrative

The previous medwatch report was submitted by william cook europe under manufacturer report reference# (b)(4). Initial reporter occupation - non-healthcare professional. Additional information provided determined that this device was manufactured by cook inc. With the submission of this initial medwatch report, cook inc. Informs that all future submissions regarding this complaint will be handled under the manufacturer report reference # referenced in this initial medwatch report. Investigation: the reported allegations have been further investigated based on the information provided to date. The following allegations have been investigated: vena cava (vc) perforation, chest pain, dyspnea. Filter interacts with ivc wall, e. G. Penetration/perforation/embedment. This may be either symptomatic or asymptomatic. Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e. G. , a surgical or endovascular procedure in the vicinity of a filter). Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration. Unknown if the reported chest pain and dyspnea are directly related to the filter and unable to identify a corresponding failure mode at this point in time. (b)(4) devices in lot. No relevant notes on work order. The product is manufactured and inspected according to specifications. No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e. G. Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava. Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c. F. R. 803. 56. This report includes information known at this time. A follow-up medwatch report will be submitted if additional relevant information become available.

 
Event Description

Patient allegedly received a cook tulip filter implant on (b)(6) 2011 via the right internal jugular vein due to deep vein thrombosis (dvt) and pulmonary embolism (pe). Patient is alleging vena cava perforation. The patient further alleges chest pain and shortness of breath. Report from ct: "there is an ivc filter in place with the filter tip at the level of the renal veins. No significant tilt is seen. The struts show preparation through the wall of the ivc into the retroperitoneal fat by approximately 1. 2 cm. There is no evidence of fracture of the struts. There is no stenosls of the inferior vena cava. ".

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameGUNTHER TULIP JUGULAR VENA CAVA FILTER SET
Type of DeviceDTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key12387951
MDR Text Key269415208
Report Number1820334-2021-02083
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK072240
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER,OTHER
Reporter Occupation
Type of Report Initial
Report Date 08/30/2021
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received08/30/2021
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device EXPIRATION Date08/01/2014
Device MODEL NumberG33017
Device Catalogue NumberIGTCFS-65-JUG
Device LOT Number2809064
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/10/2021
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured08/24/2011
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 08/30/2021 Patient Sequence Number: 1
-
-