Model Number SC-2218-50
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2021
Event Type  malfunction  
Manufacturer Narrative
Date of event: exact date unknown, event occured in (b)(6) 2021 based on the date the manufacturer became aware of the event.Sc-2218-50 (sn: (b)(4), the returned lead was analyzed, passed all tests performed, and exhibited normal device characteristics.Sc-4318 (batch/lot number: 26575414), the returned clik anchor was analyzed, passed all tests performed, and exhibited normal device characteristics.Additional suspect medical device components involved in the event: product family: scs-linear leads, upn: m365sc2218500, model: sc-2218-50, serial: (b)(4), batch: 7085396.Product family: scs-lead fixation, upn: m365sc43180, model: sc-4318, batch: 26575414.
 
Event Description
It was reported that patients leads and clik anchor were cut or damaged.The patient underwent an explant procedure.No further information has been obtained despite good faith efforts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR ST
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key12388120
MDR Text Key268879656
Report Number3006630150-2021-04857
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729767725
UDI-Public08714729767725
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/13/2022
Device Model NumberSC-2218-50
Device Catalogue NumberSC-2218-50
Device Lot Number7085338
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/10/2021
Initial Date FDA Received08/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age59 YR
-
-