| Model Number |
1012453-08 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation of a 4x8mm nc trek balloon, it was noted that there was returning excess air bubbles.The balloon kept pulling air, thus a leak was suspected.The device was not used and there was no patient involvement.Another trek balloon was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
A visual inspection was performed on the returned device.The reported leak could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during preparation of a 4x8mm nc trek balloon, it was noted that there were returning excess air bubbles.The balloon kept pulling air, thus a leak was suspected.The device was not used and there was no patient involvement.Another trek balloon was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.Subsequent to the initially filed mdr, device analysis noted a hypotube separation.The account confirmed this was not noted during preparation.No additional information was provided.
|
| |
|