ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012453-08
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/11/2021
Event Type  malfunction  
Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported that during preparation of a 4x8mm nc trek balloon, it was noted that there was returning excess air bubbles.The balloon kept pulling air, thus a leak was suspected.The device was not used and there was no patient involvement.Another trek balloon was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
A visual inspection was performed on the returned device.The reported leak could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation was unable to determine a conclusive cause for the reported complaint.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event Description
It was reported that during preparation of a 4x8mm nc trek balloon, it was noted that there were returning excess air bubbles.The balloon kept pulling air, thus a leak was suspected.The device was not used and there was no patient involvement.Another trek balloon was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.Subsequent to the initially filed mdr, device analysis noted a hypotube separation.The account confirmed this was not noted during preparation.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key12388335
MDR Text Key269001865
Report Number2024168-2021-07643
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648152054
UDI-Public08717648152054
Combination Product (y/n)N
PMA/PMN Number
K103153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2023
Device Model Number1012453-08
Device Catalogue Number1012453-08
Device Lot Number10127G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/10/2021
Initial Date Manufacturer Received 08/17/2021
Initial Date FDA Received08/30/2021
Supplement Dates Manufacturer Received09/30/2021
Supplement Dates FDA Received10/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-