| Model Number |
SN6AT3 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Headache (1880); Intraocular Pressure Increased (1937); Visual Impairment (2138); Eye Pain (4467)
|
| Date of Event |
08/11/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A optician reported that after one day of an intraocular lens (iol) implant procedure the patient experienced headache, eye pain and had tension post-op 54 mmhg.Hcp performed a pressure relief via paracentesis and later explanted the iol.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information was requested and received.Patient diagnosis with eye pressure elevation and myopia.The events outcome was unchanged.Due to events unscheduled medical treatment pars plana vitrectomy was performed.As per health care professional cause of event was flat anterior chamber.Flat anterior chamber was there due to strabismus surgery at patient age 5.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for analysis.Iol case returned opened in an unknown sealed peel pouch, adhered to unknown blister tray.A significant amount of solution is dried on both surfaces of the optic and haptics.Both haptics are intact.The optic is torn/split-cut and scratched/marked-rejectable.We are unable to determine the root cause for the reported complaint "explantation due to tension ".The returned iol shows evidence of possible handling by the customer due to the presence of solution dried on both surfaces of the optic and haptics.In addition to this, all iols are 100% cosmetically inspected as per approved manufacturing procedures and the observed lens damage would not meet our current release criteria.All product and batch history records are quality reviewed prior to product release.There have been no other complaints for this lot.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|