Model Number 20E |
Device Problem
Inappropriate/Inadequate Shock/Stimulation (1574)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Physio-control evaluated the customer's device and observed that the device has logged an event code in its memory that is indicative of a device failure which could result in a partial loss of defibrillator output energy, however no defibrillation issue was observed during testing.After performing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer for use.
|
|
Event Description
|
A customer contacted physio-control to report a non-critical issue with their device.Upon inspection, by physio-control, it was observed that their device had logged an event code in its memory that is indicative of a device failure which could result in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.There was no patient use associated with the reported event.
|
|
Manufacturer Narrative
|
Physio-control replaced the therapy pcb to resolve the logged event codes.Physio-control product analysis center (pac) further evaluated the removed therapy pcb assembly and observed that the h bridge, cr30, was shorted out in multiple directions and unable to deliver therapy.The cause of the reported issue was due to a shorted h bridge on the therapy pcb assembly.
|
|
Event Description
|
A customer contacted physio-control to report a non-critical issue with their device.Upon inspection, by physio-control, it was observed that their device had logged an event code in its memory that is indicative of a device failure which could result in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.There was no patient use associated with the reported event.
|
|
Search Alerts/Recalls
|