• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SCR BIORCI-HA 7X25 8MM HEAD STERILE; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SCR BIORCI-HA 7X25 8MM HEAD STERILE; SCREW, FIXATION, BONE Back to Search Results
Model Number 7207675
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that during the surgery, 3 biorci screws broke off when they were inserted.The procedure was performed under tourniquet.The intervention was delayed.The graft was not able to be fixed.No other complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A review of the raw material found that the storage requirements, material specifications, and applicable tests were appropriately specified.No clinically relevant information was provided for review.If retained the biosure screw remains retained in the patient, it is made of a biocomposite material which is intended for implantation to allow for bone ingrowth.If the screw was retained, it is unknown if the biosure screw will migrate and there is potential risk for tissue inflammation and/or pain.No further clinical/medical assessment is warranted at this time.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCR BIORCI-HA 7X25 8MM HEAD STERILE
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin 
5123913905
MDR Report Key12389390
Report Number1219602-2021-01814
Device Sequence Number5087861
Product Code HWC
UDI-Device Identifier03596010459930
UDI-Public03596010459930
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K002274
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/26/2025
Device Model Number7207675
Device Catalogue Number7207675
Device Lot Number50891932
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 08/11/2021
Supplement Date Received by Manufacturer08/11/2021
11/04/2021
Initial Report FDA Received Date08/30/2021
Supplement Report FDA Received Date08/30/2021
11/11/2021
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/26/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-