| Model Number |
M3535A |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/09/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to philips that the battery malfunctioned.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
It was reported to philips that the battery malfunctioned.The customer worked with the technical support representative.The customer confirmed the need for a battery exchange.The reported problem was verified in the lab.The battery pack was unable to charge either in heartstart mrx device or a cadex charger.Troubleshooting conducted revealed fuse f1 open.The fuse was replaced and the battery operated as normal.The customer was sent a battery replacement.The battery is customer replaceable.No further action at this time.
|
| |
|