• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO JOEYPUMP W/POLE CLMP; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO JOEYPUMP W/POLE CLMP; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 383400
Medical Device Problem Code Excess Flow or Over-Infusion (1311)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The incident device has been requested but to date has not been received for evaluation.  if the device is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.  as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
 
Event or Problem Description
The customer reported that the kangaroo joey enteral feeding pump overfilled by 15+ml.There was no patient harm.
 
Additional Manufacturer Narrative
An evaluation of the kangaroo pump was performed.The unit was triaged and the reported issue was confirmed.After review, it was determined that the flex cable was damaged due to misuse.Also, the software needed to be reinstalled.The flex cable was replaced.The software was reinstalled.The gasket and tie were also replaced due to opening the unit.The pump was tested and the problem was resolved.The pump is working as intended.No further actions are required at this time.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEYPUMP W/POLE CLMP
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
MDR Report Key12391324
Report Number1282497-2021-10579
Device Sequence Number4081752
Product Code LZH
UDI-Device Identifier10884521006645
UDI-Public10884521006645
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 08/11/2021
Supplement Date Received by Manufacturer08/11/2021
Initial Report FDA Received Date08/31/2021
Supplement Report FDA Received Date10/14/2021
Patient Sequence Number1
-
-