MEDTRONIC IRELAND VALIANT CAPTIVIA - FF; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number VAMF3232C100TU
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Low Blood Pressure/ Hypotension (1914); Perforation of Vessels (2135)
Date of Event 08/16/2021
Type of Reportable Event Death
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A valiant captivia stent graft was implanted in the endovascular treatment of a 46 mm thoracic aortic aneurysm.It was reported that during the index procedure, the patient's blood pressure started to drop during removal of the delivery system.The balloon was put up again to gain control.Patient expired at the end of procedure.It was stated that the external iliac tore during removal of the delivery system.As per the physician, cause of the event was patient's anatomy due to a very narrow and calcified lesion.No additional clinical sequelae were reported and the patient is expired.
 
Additional Manufacturer Narrative
Product analysis the reported vessel perforation and cause of death could not be confirmed on the films provided.Procedural angiograms during implant were not available for review.However, the anatomical considerations that were likely a factor in the reported event could be appreciated on the images.It appears likely that the patient¿s calcified and narrow vascular anatomy likely contributed to the event.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT CAPTIVIA - FF
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
MDR Report Key12391655
Report Number9612164-2021-03356
Device Sequence Number19995744
Product Code MIH
UDI-Device Identifier00643169347076
UDI-Public00643169347076
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/02/2023
Device Model NumberVAMF3232C100TU
Device Catalogue NumberVAMF3232C100TU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2021
Initial Date Received by Manufacturer 08/16/2021
Supplement Date Received by Manufacturer10/04/2021
Initial Report FDA Received Date08/31/2021
Supplement Report FDA Received Date11/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age70 YR
-
-